Iseg HV Crate and Modules

Web interface control
Normal settings
  • All channels in modules 5 to 9 should be on and set to 500 V.
  • When the HV boards are set to 0, typical currents are around 850 nA.
  • During operation (HV on and beam on), currents around 2 µA.
Possible Issues
  • Channel trip due to high current
  • Interlock signal off due to failure in LV power supply

In both cases the whole crate needs to be restarted to resume operation:

  1. Switch the crate off and on by clicking on the “Main Switch” on the picture.
  2. Run the static settings script

kfjadslj