prima_control:iseg_hv
Iseg HV Crate and Modules
Web interface control
- To select the channels corresponding to a specific submodule, click on the detector map.
Normal settings
- All channels in modules 5 to 9 should be on and set to 500 V.
- When the HV boards are set to 0, typical currents are around 850 nA.
- During operation (HV on and beam on), currents around 2 µA.
Possible Issues
- Channel trip due to high current
- Interlock signal off due to failure in LV power supply
In both cases the whole crate needs to be restarted to resume operation:
1. Switch the crate off and on by clicking on the “Main Switch” on the picture.
2. Go to the bottom right terminal window on workspace APFEL-HV
agmaas@hadanapc2:~/repo/prima-web-interface/src/iseg_hv$
3. Run the static settings script
python3 iseg_static_settings.py