==== Iseg HV Crate and Modules ==== == Web interface control == * Link: ''http://hadanapc2:8081/dcs/iseg-hv'' * To select the channels corresponding to a specific submodule, click on the detector map. == Normal settings == * All channels in modules 5 to 9 should be on and set to 500 V. * When the HV boards are set to 0, typical currents are around 850 nA. * During operation (HV on and beam on), currents around 2 µA. == Possible Issues == * Channel trip due to high current * Interlock signal off due to failure in LV power supply In both cases the whole crate needs to be restarted to resume operation: 1. Switch the crate off and on by clicking on the "Main Switch" on the picture. 2. Go to the bottom right terminal window on workspace ''APFEL-HV'' agmaas@hadanapc2:~/repo/prima-web-interface/src/iseg_hv$ 3. Run the static settings script python3 iseg_static_settings.py