* Link: '' | * Link: '' | ||
- | * All channels in modules 5 to 9 should be on and set to 500 V. | ||
* To select the channels corresponding to a specific submodule, click on the detector map. | * To select the channels corresponding to a specific submodule, click on the detector map. | ||
- | == | + | == Normal settings == |
+ | * All channels in modules 5 to 9 should be on and set to 500 V. | ||
+ | * When the HV boards are set to 0, typical currents are around 850 nA. | ||
+ | * During operation (HV on and beam on), currents around 2 µA. | ||
+ | |||
+ | == Possible Issues == | ||
+ | * Channel trip due to high current | ||
+ | * Interlock signal off due to failure in LV power supply | ||
+ | |||
+ | In both cases the whole crate needs to be restarted to resume operation: | ||
+ | |||
+ | 1. Switch the crate off and on by clicking on the "Main Switch" | ||
+ | |||
+ | 2. Go to the bottom right terminal window on workspace '' | ||
+ | |||
+ | | ||
+ | |||
+ | 3. Run the static settings script | ||
+ | |||
+ | | ||
+ | |||
+ |