-  * ramp HV up: + 
 +=== Ramp detector HV === 
 + 
 + 
 +== ramp Detector HV up: ==  
 +1) check that HV on ISEG is ramped up to 500V; IF NOT, RAMP UP (SEE BELOW) 
 + 
 +2) ramp up detector HV: 
 + 
 +in workspace "Apfel+HV" find terminal Ramp HV:
 <code> <code>
-agmaas@prima-daq1:~/repo/hv_control$ python3 ramp_hv.py -f /data/data_raid/prima_feb_25/hv_scan_6 -x /home/agmaas/Data/beamtime_Feb2025/hv/exclude_channels.dat -id all -off -0+agmaas@prima-daq1:~/repo/hv_control$ python3 ramp_hv.py -f /data/data_raid/prima_may_25/hv_scan_2/ -x /home/agmaas/Data/beamtime_May2025/hv/exclude_channels.dat -id all -off 0
 </code> </code>
 +
 +2b) if half of submodule 32 needs to be ramped down, use the alternative exclude file:
 +
 +<code>
 +agmaas@prima-daq1:~/repo/hv_control$ python3 ramp_hv.py -f /data/data_raid/prima_may_25/hv_scan_2/ -x /home/agmaas/Data/beamtime_May2025/hv/exclude_channels_sub32high_off.dat -id all -off 0
 +</code>
 +
 +
 +== ramp Detector HV down: ==
 +
 +1) ramp down HV on APDs
 +<code>
 +agmaas@prima-daq1:~/repo/hv_control$ python3 ramp_hv.py -val 0 -id all
 +</code>
 +if the last line of the output is a 1 repeatedly, you can stop the terminal "Crt+c"
 +
 +
 +**--------------------------------------------------------------------------------------------------------------------------------------------------**
 +
 +
 +=== Ramp ISEG HV ===
 +
 +== ramp pre-HV (ISEG) up: ==
 +
 +1) **check that detector hv is off**
 +
 +2) use web interface (tab should be open in firefox already)
 +
 +IF NOT: http://hadanapc2:8081/dcs/iseg-hv
 +
 +3) for modules 5 to 9, click checkbox "select all" and click "switch all <on/off>"
 +
 +
 +== ramp pre-HV (ISEG) down: ==
 +
 +1) **check that detector hv is off**
 +
 +2) use web interface (tab should be open in firefox already)
 +
 +IF NOT: http://hadanapc2:8081/dcs/iseg-hv
 +
 +3) for modules 5 to 9, click checkbox "select all" and click "switch all <on/off>"
 +
  
  
 +  
  
 +{{:prima_control:screenshot_2025-02-26_13-08-57.png?600|}}
  
  
  